ZR-3211H UV DOAS Method GAS Analyzer

Short Description:

ZR 3211H  Stack dust( gas)  Tester by UV Differential Optical absorption spectroscopy is a portable instrument,which can measure the concentration of SO2, NOx, O2, NH3. It cannot works affected by the water vapor in the flue gas,which is especially suitable for high humidity and low sulfur conditions. The instrument is designed to be integrated with the host and convenient for sampling. It can be used by environmental departments to test the gas concentration and emission from boilers, and can also be used in industrial and mining enterprises to measure the concentration of various harmful gases.


Product Detail

Specification

ZR 3211H  Stack dust( gas)  Tester by UV Differential Optical absorption spectroscopy is a portable instrument,which can measure the concentration of SO2, NOx, O2, NH3. It cannot works affected by the water vapor in the flue gas,which is especially suitable for high humidity and low sulfur conditions. The instrument is designed to be integrated with the host and convenient for sampling. It can be used by environmental departments to test the gas concentration and emission from boilers, and can also be used in industrial and mining enterprises to measure the concentration of various harmful gases.

Applications

>  Measure the O2 of gas and excess air coefficient.

>  Evaluate and calibrate the accuracy of continuous flue gas measuring instruments.

>  Other applicable occasions.

Standards

>  GB13233-2011     Fossil-fuel power plant atmospheric pollutant emission standard

>  GB/T37186-2018 Gas analysis -- Determination of sulfur dioxide and nitrogen oxides -- ultraviolet differential absorption spectrometric method

>  HJ 973-2018        Determination of carbon monoxide in exhaust gas from fixed pollution sources  constant potential electrolysis

>  HJ/T 397-2007     Technical specification for fixed source waste gas monitoring

>  HJ 1045-2019      Technical requirements and detection methods of portable ultraviolet absorption measuring instruments for flue gas (SO2 and NOx) from fixed pollution sources

>  JJG 968-2002      Verification regulation of flue gas analyzer

Features

>  Designed to be integrated with the host and convenient for sampling.

>  Equipped with titanium alloy vacuum heat insulation pipe, good heat insulation effect.

> Tested by UV Differential Optical absorption spectroscopy. It can measure the concentration of SO2, NOx, NH3 and cannot works affected by the water vapor in the flue gas,which is especially suitable for high humidity and low sulfur conditions.

>  Automatically switch the range according to the high and low concentration values of SO2, NO and NO2.

> Equipped with pitot tube and smoke temperature sensor, it can measure smoke temperature flow rate and moisture content.

> Self contained battery≥3H.

> Dynamically save sampling data and spectral data, and export excel table.

> Built in condensate water removal module to prevent water from entering the sensor

 Working conditions

> Power supply:AC220V±10%,50Hz or DC24V 12A

> Ambient temperature:(-20~ 45 )℃

> Ambient humidity:0% ~95%

> Atmospheric pressure: (60~130) kPa

> Application environment: Non-explosionproof 

> When used in wild,some measurement should be adopt to prevent the damage caused by rain,snow,dust and sunshine.

> Good power grounding

Deliver Goods

deliver goods Italy
  • Previous:
  • Next:

  • Host parameters

    Parameter Range Resolution Error
    Flue gas temperature (0~200)℃ 1℃ ±3.0℃
    Flue gas static pressure (-30~30)kPa 0.01kPa ±2.0%FS
    Flue gas dynamic pressure (0~2000)kPa 0.01kPa ±2.0%FS
    Moisture content (0~40)VOL% 0.01VOL% ±2.0%
    Sampling flowrate ≥0.5L/min 0.1L/min ±2.5%
    Total pressure (60~130)kPa 0.01kPa ±0.5kPa
    Excess air coefficient 1~99.99 0.01 ±2.5%
    Working temperature (-20~50)℃
    Load capacity of sampling pump ≥40kPa
    Data storage >1000000 groups
    Power Supply AC220V±10%,50Hz
    Size (length 1270×width 120×height 248)mm
    Weight About 5.5kg(battery include)
    Power consumption <120W

    Flue gas sampling parameters

    Parameter

    Range

    Resolution

    Error

    SO2

    Low range:(0~430)mg/m3

    High range:(0~5720)mg/m3

    0.1 mg/m3

    Relative error:± 3%Repeatability:≤1.5%Response time:≤90sStability:Indication change within 1h<5%Detection limit:

    SO2≤2mg/m³

    NO≤1mg/m³

    NO2≤2mg/m³

    NO

    Low range:(0~200)mg/m3

    High range:(0~1340)mg/m3

    0.1 mg/m3

    NO2

    Low range:(0~300)mg/m3

    High range:(0~1000)mg/m3

    0.1 mg/m3

    NH3

    (0~300)mg/m3

    0.1 mg/m3

     

    Write your message here and send it to us